Making Physiology Happen

Products

IX-100BE Data Acquisition System

Description

The iWorx IX-100BE Data Acquisition System provides high resolution/low noise recording of biopotential signals for OEM applications. With one isolated input channel and a low voltage stimulator, the system is ideal for embedded biomedical and life science applications where only one biopotential recording channel is required.

The iWorx IX-100BE has one isolated input channel and is equipped with a low voltage stimulator and an audio output.

Features

Resolution and Noise
The iWorx IX-100BE uses a 16-bit A/D converter to sample data over its full input range of ±1V at speeds up to 200 kHz. Its low noise greatly reduces the need for gain and offset.

Versatile Bioamplifier
The iWorx IX-100BE offers a wide selection of low pass and high pass filters* and is safe for human connection.

*IX-100BE high pass filter values (MHz): 0.03, 0.3, 3, 30, 100

*IX-100BE low pass filter values (MHz): 10, 20, 30, 40, 50, 60, 70, 80, 90, 100, 150, 200, 300, 400, 500, 600, 700, 800, 900, 1000, 2000, 3000, 4000, 5000, 6000, 7000, 8000, 9000, 10000, 20000

Input Range
±1V, ±250mV, ±125mV, ±50mV, ±25mV, ±12.5mV, ±5mV, ±2.5mV

Low Voltage Stimulator
The iWorx IX-100BE has a software programmable 16 bit, +/-5V stimulator (DAC). Parameters for stimulator, such as pulse width, frequency and amplitude, may be changed “on-the-fly” using handy controls located in the LabScribe2™ software tool bar. Standard protocols include Pulse, Train, Step, Triangle and Ramp. In applications involving isolated
heart, the stimulator may be programmed for pacing.

Software
The iWorx IX-100BE recorder is controlled by LabScribe2 software, a powerful recording and analysis software package. LabScribe2 has an intuitive, user-friendly interface for setting up acquisition screens, calibrating signals and analyzing data. Up to 128 channels of data can be displayed simultaneously at sampling rates as high as 100,000
samples/second.

A comprehensive set of analysis routines have been pre-configured making data analysis and interpretation quick and easy.

Specifications

iWorx IX-100BE Specifications
 
No. of Channels 1
Input Range Software Configurable
Resolution 16 Bit
Sampling Speed 200 kHz
High Pass Filters Software Configurable
Low Pass Filters Software Configurable
Isolation 3 kV
Connector IEC-60601-1 Touch Proof
 
General
Power 5 VDC 2.6A
Enclosure Plastic
Interface USB
Enclosure Plastic
Warranty 3 years
Software LabScribe2 Recording and Analysis Software


Part Number: IX-100BE